EquipmentKosmos 330 Diagnostic X-ray system

A stationary single workstation diagnostic X-ray system, designed for chest imaging in vertical position. A mobile table version allows for X-ray examinations in supine and prone positions. A choice of options allows the unit to be adapted to the needs of a healthcare facility.

The imaging system, based on flat-panel detectors, provides diagnostic-quality images with reduced radiation exposure.

Typical configuration:

Item Quantity, unit
X-ray power supply unit 1
X-ray tube 1
Floor-mounted stand 1
Wall stand 1
X-ray collimator 1
Anti-scatter grid 1
Ionization chamber 1
Digital radiography image acquisition system 1
Dose-area product (DAP) meter 1
Main electrical specifications
X-ray power supply device
Power supply Three-phase 400 V ±10%, 50 Hz
Maximum generator power, kW ≥ 32 *
Inverter frequency, kHz ≥ 25
Anode voltage, kV 40 – 125 (in 1 kV step)
Anode current, mA 10 – 500 *
Exposure time, s 0,001 – 6,3
X-ray tube
Nominal voltage, kV 125 *
Dual focal spots, mm 0,6 / 1,2 *
Anode heat capacity, kHU 300 *
Anode angle, degrees 12,5
Maximum leakage radiation, mGy/h < 0.44 at 1 m

* Specifications may be modified upon customer request.

Main technical specifications
Flat-panel detector
Flat-panel location Wall stand
Scintillator Cesium Iodide (CsI)
Field of view, cm 43 х 43
Image matrix, pixels 3072 х 3072
Pixel size, μm 140
Spatial resolution, lp/mm ≥ 3,5
Image acquisition time, s ≤ 2
Grayscale, bit 16
Installation type Built-in
Image transmission Wired (Ethernet)
Quantity, unit 1
Main functional specifications
Digital radiography image acquisition system
Patient data entry, image acquisition and preview, preliminary image processing
X-ray power supply unit control function
Detector automatic synchronization at the start of exposure
Anatomically programmed radiography exposure parameters transmission to the generator and selection of workstation
Option to install additional flat-panel detectors (total number of detectors — up to 4
units)
Preparation for examination
Patient data entry • DICOM worklist server query
• Search in local database
• Manual data entry via keyboard
• «Emergency» mode
Exposure parameters automatic setting (when selecting anatomical program) • Exposure mode selection (one-point / two-point technique, AEC mode)
• Exposure parameters
• Image processing algorithms
Image viewing and processing
Study search using various filters
Full-screen and mosaic image viewing
Rejection of selected images
Image post-processing functions • Vertical and / or horizontal image flipping
• 90° image rotation
• Automatic or manual (mouse-controlled) brightness and contrast (window width / level) adjustment
• Grayscale inversion (positive / negative)
• Image filters (edge enhancement, noise reduction, contrast enhancement, edge sharpening, etc.)
• Annotation tools (text, radiological markers)
• Distance and angle measurements
• Measurement calibration
• Image cropping (electronic collimation)
• Zoom
• Stitching function
Automatic image transfer to DICOM devices upon study completion
Background image archiving with continued operation at all system workstations
Image storage
Storage capacity: not less than 1500 images
DICOM compatibility
DICOM services • Storage SCU
• Print SCU
• MWL SCU
• CD/DVD media export (includes built-in viewer software compatible with Windows OS)