EquipmentKosmos Universal/A Digital diagnostic X-ray system for three workstations

Digital diagnostic X-ray system for three workstations.

Typical configuration:

Item Quantity, unit
X-ray power generator 1
X-ray tube 2
Automatic X-ray collimator 1
Manual X-ray collimator 1
Floor-mounted tube stand 1
Anti-scatter grid 3
Ionization chamber (AEC sensor) 3
Mobile patient table 1
Wall stand 1
Tilting X-ray table 1
Digital radiography & fluoroscopy system 1 set
Dose Area Product (DAP) meter 1
Main specifications
X-ray generator
Power supply Three-phase 400 V ±10%, 50 Hz
Nominal power, kW 50
Maximum power consumption, kVA 65
Inverter frequency, kHz 100
Tube Current (Fluoroscopy Mode), mA Continuous: 0,5 – 10
Pulsed: 5 – 99
Tube Voltage (Radiography Mode), kV 40 – 150 (1 kV increments)
Tube Voltage (Fluoroscopy Mode), kV 40 – 125 (1 kV increments)
Automatic brightness control system for fluoroscopy
X-ray tube assembly
Nominal tube voltage, kV 150 *
Focal spot size, mm 0,6 / 1,2
Anode heat capacity, kHU 300
Anode target angle, degrees 12,5
Maximum leakage radiation, mGy/h < 0,44 at 1 m

* Specifications may be modified upon customer request.

Main technical specifications
Tilting table system
Table tilting angle (motorized), degrees from −15 to +90
Longitudinal motorized table movement, mm:
• toward patient foot
• toward patient head
• 600
• 400
Transversal motorized table top movement, mm 200
Longitudinal manual & motorized movement SFD / tube, mm ≥ 500
Transversal manual movement of imaging assembly, mm ≥ ±100
Vertical manual movement SFD movement, mm ≥ 195 – 460 (source-to-image distance 670–920)
Combined motorized movement of table top and SFD / tube synchronized tracking for contrast media procedures
Controls (Integrated Contro Panel) • Motorized movements of table, and tilting for SFD
• X-ray field collimation
• Vertical and horizontal image flipping
• Magnification mode selection (3 field of view)
• Compressor movement activation
• Electromagnetic brakes activation / deactivation
Main functions
Digital fluoroscopy / radiography system
Detector installation type Built-in
Scintillator Cesium Iodide (CsI)
Field of view 12” х 12”
Active imaging area, mm 297 х 297
Spatial resolution, lp/mm • 1.7 at 12″ x 12″ (no magnification)
• 1.7 at 9″ x 9″ (magnification 1)
• 3.4 at 6″ x 6″ (magnification 2)
Spatial resolution in radiography mode, lp/mm 3,4
Image matrix, pixels: • Fluoroscopy
• Radiography
• 1024 х 1024
• 2048 х 204
Image acquisition modes
Mode Matrix, pixels Frame rate, fps
Continuous fluoroscopy 1024 x 1024 25
Pulsed fluoroscopy 1024 x 1024 3; 7,5; 15
Radiography 2048 x 2048 (no magnification)
1536 x 1536 (magnification 1)
1024 x 1024 (magnification 2)
Serial radiography (up to 10 s, user-defined) 1024 x 1024 (no magnification)
768 x 768 (magnification 1)
512 x 512 (magnification 2)
3